H60 Heated Humidifier User Manual V2.0
3 / 21
WARNINGS!
• Use the humidifier only for its intended use as described in this manual.
• Use only accessories recommended by BMC.
• Do not bring the device or accessories into a Magnetic Resonance (MR) environment as it
may cause unacceptable risk to the patient or damage to the device or MR medical devices.
The device and accessories have not been evaluated for safety in an MR environment.
• Do not use the device or accessories in an environment with electromagnetic equipment
such as CT scanners, Diathermy, RFID and electromagnetic security systems (metal
detectors) as it may cause unacceptable risk to the patient or damage to the device. Some
electromagnetic sources may not be apparent, if you notice any unexplained changes in the
performance of this device, if it is making unusual or harsh sounds, disconnect the power
cord and discontinue use. Contact your home care provider.
4.1 Connecting, Separating the Humidifier and Main
Device
4.1.1 Connecting the Humidifier to the Main Device
Remove the shield from the main device, following the steps below:
(1) Overturn the main device and find the buckle slot at the bottom of the main device, as
shown in Fig. 4-1.
(2) Remove the shield by inserting a flat tool into the buckle slot.
Fig. 4-1
After the shield is removed, place the humidifier and main device near each other as shown
in Fig. 4-2. The air outlet of the main device should be targeted to the inlet of the humidifier.
Push the two devices together until they click into place. Fig. 4-2 shows their positions before
and after connection to each other.
Before Connection After Connection
Fig. 4-2