Achieving a good image13
2. Select
(Temperature scale) and push the navigation pad. This displays a submenu.
3. Select
(Manual) and push the navigation pad.
4. To simultaneously change the temperature scale minimum and maximum limits, push
the navigation pad up/down.
5. To change the minimum limit or the maximum limit, do the following:
• Push the navigation pad left/right to select (highlight) the maximum or minimum
temperature.
• Push the navigation pad up/down to change the value of the highlighted
temperature.
13.4 Changing the camera temperature range
13.4.1 General
The camera is calibrated for different temperature ranges. Available temperature range
options are dependent on the camera model.
For accurate temperature measurements, you must change the Camera temperature
range setting to suit the expected temperature of the object you are inspecting.
Note For more information, see section 33 About calibration, page 233.
13.4.2 Procedure
Follow this procedure:
1. Push the navigation pad to display the menu system.
2. Select
(Settings) and push the navigation pad. This displays the Settings menu.
3. Select Camera temperature range and push the navigation pad. This displays a dialog
box.
4. Select the appropriate temperature range and push the navigation pad.
Note You can also assign the function Switch temperature range to the programmable
button
. For more information, see section 10.12 Assigning functions to the program-
mable button, page 37.
#T810190; r. AI/41890/41890; en-US
62