DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
8 - 412 Section 8-16 - Options replacement
Calibration and adjustments
No calibrations or adjustments are needed after this part replacement.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws that you removed earlier have been installed.
2.) Connect cables and probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional tests to confirm the system is operational before returning the system
to the customer.
Table 8-406 4D Option replacement Checks
See
Section
Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
5750007-1EN. Equipment passed all required
tests and is ready for use.
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks
Perform a 4D sweep
4-2-4 Power SHUT DOWN