Installation
2030047-001A Patient Data Module 3-5
Connect to host patient monitor
Connect to transport bedside monitor
If using the Patient Data Module with a transport bedside monitor,
1. Guide the mounting rails of the Patient Data Module onto the transport dock
mounting rails.
2. Slide the Patient Data Module toward the ePort until the locking key secures it to
the transport dock.
Connect to bedside monitor
If using the Patient Data Module with a bedside monitor,
1. Connect one end of the ePort host interface cable to the Patient Data Module
ePort connector.
2. Connect the other end of the ePort host interface cable to the Solar to Patient
Data Module adapter on the back of the Solar 8000M/i patient monitor.