4-10 Patient Data Module 2030047-001A
Service Tool and Configuration
Diagnostics
The Diagnostics category consists of Log Files and Service Metrics modules.
Log Files
The Log Files module allows viewing and downloading log files. Messages and
errors in log files provide useful information to a trained technician.
Click View Log to view or Download Log to download the log file.
Service Metrics
The Service Metrics module displays various Patient Data Module statistics. NBP
STATISTICS can be useful in determining NBP usage.
Click REFRESH ALL to refresh the data.
The following buttons clear NBP data:
NBP Pump Run time
NBP Deflate Valve Cycle Count
NBP Dump Valve Cycle Count
NBP Pump Cycle Count
NOTE
Some Patient Data Module statistics only update every 30 seconds.
After replacing the NBP pump assembly, clear all NBP statistics and perform the
checkout procedures provided in the host patient monitor service manual.
If a value in the Service Metrics module is out of specification, the data field will
be highlighted red. Contact GE technical support. See the “How to Reach Us...”
page included with this manual for contact information.
Calibration
The Calibration category consists of Analog-Out ECG Calibration and Analog-Out
IP Calibration. The following additional equipment in required to perform this
calibration:
Unterminated defib sync cable (2017842-001, part of the Patient Data Module
Service Tool Kit)
Digital voltmeter
1. Connect the unterminated cable to the Defib/ Sync connector on the front of the
Patient Data Module.
2. Click Start Calibration to begin the calibration process.
3. Follow instructions displayed on screen.
NOTE
To abort the calibration process with no changes, press Stop Calibration.