A-2 Patient Data Module 2030047-001A
Appendix A–Technical Specifications
Technical Specifications
Physical
Environmental
Battery
Height: 7.0 cm (2.75 in)
Width: 14.6 cm (5.75 in)
Depth: 21.6 cm (8.5 in)
Weight: 1.1 kg (2.5 lbs.) without optional battery
1.3 kg (2.8 lbs.) with optional battery
Power requirements
Input voltage:
Input current:
8 to 17 Vdc
0.3A nominal at 16.75 V
Power consumption: 4.5 Watts, nominal
Cooling: Natural convection
Operating heat dissipation: 10.2 Btu/Hr
NOTE
System may not meet its performance specifications if stored or used outside the
manufacturer’s specified temperature and humidity ranges.
Operating temperature and
humidity:
10º C to 35º C (50º F to 95º F)
15% to 95% RH non-condensing
Storage temperature and
humidity:
-40º C to 60º C (-40º F to 140º F)
15% to 95% RH non-condensing
Type: Exchangeable lithium ion
Quantity: One
Voltage: 11.1 V nominal
Capacity: 1.8 Amp hour nominal
Run time: Approximately 3.5 hours with new fully charged battery
Charge time: Approximately 2 hours
Battery life: 500 cycles to 50% capacity