A-10 Patient Data Module 2030047-001A
Appendix A–Technical Specifications
Cardiac output
Certifications
Method: Thermodilution
Parameters Displayed: Cardiac output, blood temperature, injectate temperature,
real-time cardiac output washout curve, last average CO
Cardiac output range: 0.2 to 15 liters per minute
Blood temperature range: 17°C to 42°C (62°F to 107°F)
Blood temperature
accuracy:
±0.5°C 17°C to 30°C
Injectate temperature
range:
0°C to 30°C (32°F to 86°F)
Injectate temperature
accuracy:
±0.3°C
Cardiac output review: Accept/reject individual measurements and store average
Catheter sizes: 5, 6, 7, 7.5, or 8 French
Injuectate volume
selections:
3, 5, or 10
IEC/EN/UL 60601-1
CAN/CSA C22.2 No. 601.1
IEC/EN 60601-1-2
IEC/EN 60601-2-27
IEC/EN 60601-2-30
IEC/EN 60601-2-34
IEC/EN 60601-2-49
IEC/EN 60601-2-51
EN 12470-4
EN ISO 9919
ANSI/AAMI SP10
ANSI/AAMI EC11
ANSI/AAMI EC13
CE Marking: Medical Devices Directive - 93/42/EEC
JJG 760
YY 91079