GE Healthcare Senographe DS Acquisition System
Revision 1 Operator Manual 5307907-3-S-1EN
Image Acquisition Procedure
12-imgac.fm Page no. 127 Chapter 12
4 AOP - Automatic Optimization of Parameters
4-1 Definition of AOP
AOP (Automatic Optimization of Parameters) automates radiation controls by optimizing the exposure
parameters.
It can be used for contact exposures, magnified exposures, or for the examination of the chest wall. See
section 3 .Configuration on page 126 for additional notes on configuration.
Before each exposure made in AOP mode, a brief absorption measurement exposure is made to deter-
mine the thickness and density of the breast being examined. This information, together with the priority
set by the operator, allows optimization of exposure parameters.
AOP automatically selects the following parameters:
Anode track (Mo or Rh)
Filter (Mo or Rh)
kV value
mAs value
The exposure is then made, using the track/filter/kV/mAs configuration which supplies the lowest dose
with respect for the operator’s imaging priority.
4-2 Usage of AOP Mode
The operator chooses one of three modes, according to the priority required:
CNT Contrast Gives priority to subject contrast.
STD Standard Selects a compromise between contrast and dose.
DOSE Dose Gives priority to dose reduction.
Note:
By default, the system selects the large focal spot when a contact exam is selected, and the small
focal spot when magnification is selected. Exposures are not possible if these settings are overrid-
den.
Note:
GE recommends that Standard (STD) and Dose (DOSE) modes be used for routine mammogra-
phy. The Senographe DS Contrast (CNT) mode delivers more dose than Standard and Dose
modes. GE recommends that Contrast mode be used only after consultation with an interpreting
physician or radiologist.
Note:
To fulfill the requirements of certain countries, access to Contrast Mode can be restricted as de-
scribed in the Service Documentation.
CAUTION
Breast compression of at least 3 daN (30 newtons or 6.7 pounds) is essential when using
the AOP mode.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals