System Setup
3-74 Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
3-10 Option Setup
3-10-1 Software Option Installation Procedure
NOTE: Not all features described in this section may be available or
cleared for sale in all markets. Please contact with your local GE
Ultrasound representative to get the latest information.
1. Power on the system.
NOTE: Keep the power cord connection during the installation.
2. After the power-up sequence is complete, press Config (F2)
on the control panel, and then select Admin.
3. In System Admin tab, select New.
Figure 3-64. New Option Key
For software Option Installation, the operator must login as
Administrator.