EasyManuals Logo

GE Vivid T8 User Manual

GE Vivid T8
349 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #319 background imageLoading...
Page #319 background image
Tools required
Vivid T9/Vivid T8 – Basic Service Manual 10-11
5795591-100 English Rev. 3
10-5-3 Special tools, supplies and equipment used for maintenance
Table 10-4: Overview of tool requirements for periodic maintenance
Tool / kit Comments
Digital Volt Meter (DVM)
Anti Static Kit Kit includes anti–static mat, wrist strap and cables for 200
to 240V system
3M #2204 Large adjustable wrist strap
3M #2214 Small adjustable wrist strap
3M #3051 conductive ground cord
Anti Static Vacuum Cleaner 120V
230V
Safety Analyzer The safety Analyzer tool should be calibrated and
compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS
3551.
QIQ Phantom RMI Grayscale Target Model 403GS
NOTE! The use of a Phantom is not required during
Preventive Maintenance. Customer may use it as part
of their Quality Assurance Program tests.
B/W Printer Cleaning Sheet See printer user manual for requirements
Color Printer Cleaning Sheet See printer user manual for requirements
Disposable Gloves

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid T8 and is the answer not in the manual?

GE Vivid T8 Specifications

General IconGeneral
BrandGE
ModelVivid T8
CategoryMedical Equipment
LanguageEnglish

Related product manuals