Replacement Procedures
8-20 Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
8-4 Loading the software
8-4-1 Purpose of this section
This section describes how to reinstall software on Vivid T9/
Vivid T8.
8-4-2 Customer provided prerequisite
• Formatted and labelled media for Images storage.
• Formatted and labelled media for Patient Archive and
Presets (User Defined Settings).
• Password for the user ADM.
8-4-3 Data Management - moving all images
• Move the images to a remote server or to removable media.
For instructions, please see “Disk management” in the User
Manual/User Guide.
An error, or a power loss may occur.
Always backup the Patient Archive and the Presets (System
Configurations) before loading the software!
In order to complete a successful restore of the Patient
Database, as needed after a hard disk replacement, or if all the
content on the hard disk has been erased, the images must be
moved away from Vivid T9/Vivid T8 before doing backup of the
Patient Database.
Depending on the location set-up, either move the images to a
remote server or to removable media like DVD or CD discs.