8.5 Cleaning
Safety precautions
WARNING
Squeezing or crushing injuries. Due to the significant weight of
the system, great care must be taken not to cause squeezing or
crushing injuries during movement.
CAUTION
Hot surface. Make sure that the instrument surface temperature
is equal to room temperature before cleaning.
CAUTION
Cleaning. Always clean the equipment in a well ventilated area.
Never immerse any part of the equipment in liquid or spray liquid
on the equipment. Always make sure that the equipment is com-
pletely dry before plugging it in. Always follow all environmental,
health, and safety guidelines pertaining to the materials used.
Resistance to cleaning agents
XDM and XDUO mixers are resistant to the following cleaning agents:
ConcentrationCleaning agent
UndilutedActril™
Diluted 1:128Environ
®
LpH
®
70%Ethanol
UndilutedKlercide™-CR
1.5%NaClO
70%2-propanol
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD 257
8 Maintenance
8.5 Cleaning