10.7.3 TCU coolant specifications
TCU water quality specifications
Tap water can be used to prepare the coolant, when the coolant temperature (mixer
jacket temperature) is designated to be between 6°C and 65°C (the temperature of the
disposable bag contents is between 10°C and 60°C ). The water quality must be within
the guidelines shown below. Deionized water is not suitable.
Contact the TCU manufacturer for additional information.
SpecificationParameter
NoneSuspended solids
50 to 600 µS/cmConductivity
< 80 °dHHardness
7.8pH
NoneCO
2
aggressive
< 0.3 mg/LIron
< 0.05 mg/LManganese
< 250 mg/LSulfate
< 250 mg/LChloride
< 40 mg/LCOD
< 1000 CFU/mLBacteria
TCU coolant composition
The use of corrosion inhibitor in the coolant is optional. If the vessel jacket temperature
is expected to be below +6°C, antifreeze needs to be added to TCU coolant.
The table below shows an example coolant composition (containing corrosion inhibitor)
that allows the vessel jacket temperature to decrease to -15°C without freezing.
ComponentAmount (%)
OptiShield
®
Plus10
propylene glycol50
water40
288 Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
10 Reference information
10.7 Temperature control unit
10.7.3 TCU coolant specifications

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Xcellerex XDM Series and is the answer not in the manual?

GE Xcellerex XDM Series Specifications

General IconGeneral
BrandGE
ModelXcellerex XDM Series
CategoryMixer
LanguageEnglish

Related product manuals