Disabling Parallel Poll Responses
The PPU (Parallel Poll Unconfigure) statement gives the controller the
capability of disabling the parallel poll responses of one or more devices on
the bus.
Example The following statement disables device 5 only:
PPOLL UNCONFIGURE 705
This statement disables all devices on interface select code 8 from
responding to a parallel poll:
PPOLL UNCONFIGURE 8
If no primary address is specified, all bus devices are disabled from
responding to a parallel poll. If a primary address is specified, only the
specified devices (which have the parallel poll configure capability) are
disabled.
HP-IB Commands
The following paragraphs describe actual HP-IB commands which can be
used to perform the functions of the BASIC commands shown in the previous
examples.
Parallel Poll Unconfigure Command
The parallel poll unconfigure command (PPU) resets all parallel poll devices
to the idle state (unable to respond to a parallel poll).
Parallel Poll Configure Command
The parallel poll configure command (PPC) causes the addressed listener to
be configured according to the parallel poll enable secondary command PPE.
Status Reporting
Disabling Parallel Poll Responses
7–13

Table of Contents

Other manuals for HP 16500C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 16500C and is the answer not in the manual?

HP 16500C Specifications

General IconGeneral
BrandHP
Model16500C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals