Query :SYSTem:DATA?
The SYSTem:DATA query returns the block data. The data sent by the
SYSTem:DATA query reflects the configuration of the selected module when
the last acquisition was performed. Any changes made since then through
either front-panel operations or programming commands do not affect the
stored data. Since the mainframe does not acquire data, refer to the
appropriate module Programmer’s Guide for more details.
Returned Format
[:SYSTem:DATA] <block_data><NL>
Example See the Programmer’s Guide for the selected module for an example.
DSP (Display)
Command :SYSTem:DSP <string>
The DSP command writes the specified quoted string to a device-dependent
portion of the instrument display.
<string> A string of up to 68 alphanumeric characters
Example OUTPUT XXX;":SYSTEM:DSPThe message goes here’"