EasyManuals Logo
Home>HP>Measuring Instruments>16500C

HP 16500C Programmer's Guide

HP 16500C
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #179 background imageLoading...
Page #179 background image
AUToload
Command :MMEMory:AUToload {{OFF|0}|{<auto_file>}}[,<msus>]
The AUToload command controls the autoload feature which designates a set
of configuration files to be loaded automatically the next time the instrument
is turned on. The OFF parameter (or 0) disables the autoload feature. A
string parameter may be specified instead to represent the desired autoload
file. If the file is on the current drive, the autoload feature is enabled to the
specified file. The configuration files specified must reside in the root
directory of the current drive.
<auto_file> A string of up to 10 alphanumeric characters for LIF in the following form:
NNNNNNNNNN
or
A string of up to 12 alphanumeric characters for DOS in the following form:
NNNNNNNN.NNN
<msus> Mass Storage Unit specifier. INTernal0 for the hard disk drive and
INTernal1 for the flexible disk drive.
Example Disabling autoload:
OUTPUT XXX;":MMEMORY:AUTOLOAD OFF"
Setting FILE2 on the hard drive as the autoload file:
OUTPUT XXX;":MMEMORY:AUTOLOAD ’FILE2 ’,INTERNAL0"
Query :MMEMory:AUToload?
The AUToload query returns 0 if the autoload feature is disabled. If the
autoload feature is enabled, the query returns a string parameter that
specifies the current autoload file.
Returned Format
[:MMEMory:AUToload] {0|<auto_file>},<msus><NL>
Example OUTPUT XXX;":MMEMORY:AUTOLOAD?"
MMEMory Subsystem
AUToload
12–7

Table of Contents

Other manuals for HP 16500C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 16500C and is the answer not in the manual?

HP 16500C Specifications

General IconGeneral
BrandHP
Model16500C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals