If
the Green
LEDs on
A8
are
not
All
ON
The green
LEDs along
the
top
edge
of
the
A8
p
ost
regulator
are
normally
on.
Flashing
LEDs on
A8
indicate
that
the
sh
utdo
wn
circuitry
on
the
A8
post
regulator is
protecting
p
o
w
er
supplies from
ov
ercurrent
conditions b
yrep
eatedly sh
utting them
do
wn.
This ma
yb
e caused
by
supply
loading
on
A8
or
on
an
y
other
assem
bly
in
the
analyzer.
Remo
v
e
A8,
Maintain
A15W1
Cable
Connection
1. Switc
h o
the analyzer.
2.
Remo
v
e
A8
from
its
motherb
oard
connector,
but
k
eep
the
A15W1 cable
connected to
A8.
3.
Remo
v
e
the
displa
y
po
wer
cable W14.
See
Figure
5-5
.
4. Short
A8TP2 (A
GND)
(see
Figure
5-3
)
to
c
hassis
ground
with
a
clip
lead.
5.
Switc
h
on
the
analyzer
and
observ
e
the
green
LEDs
on
A8.
If
an
y
green
LEDs
other
than
+5VD
are
still
o
or
ashing, con
tin
ue
to
\Chec
k
the
A8
F
uses
and
V
oltages."
If
all
LEDs
are
no
w
on
steadily
except
for
the
+5VD
LED,
the
A15
preregulator
and
A8
p
ost regulator
are w
orking
prop
erly
and
the
trouble
is
excessiv
e
loading
somewhere
after
the
motherb
oard
connections
at
A8.
Con
tin
ue to
\Remov
e
the
Assem
blies."
Check
the
A8
Fuses
and
V
oltages
Chec
k
the
fuses
along
the
top
edge
of
A8.
If
an
y
A8
fuse
has
burned
out,
replace
it. If
it
burns
out
again when
p
o
w
er
is
applied
to
the
analyzer,
A8
or
A15
is
fault
y
.
Determine
whic
h
assem
bly
has
failed
as
follo
ws:
1.
Remo
v
e
the
A15W1
cable
at
A8.
See
Figure
5-5
.
2. Measure
the v
oltages
at
A15W1P1
(see
Figure
5-6
)
with
a
v
oltmeter
ha
ving
a
small
prob
e.
3.
Compare the
measured
v
oltages
with
those
in
T
able
5-2
.
If
the
v
oltages
are within
tolerance, replace
A8.
If
the
v
oltages
are
not
within
tolerance,
replace
A15.
4.
If
the
green LEDs
are
no
w
on,
the
A15
preregulator
and
A8
p
ost regulator
are
w
orking
properly and the trouble is excessiv
e loading somewhere after the motherb oard connections
at A8. Con
tinue to \Remo
ve
the Assem
blies."
Remove the Assemblies
1.
Switch o the analyzer.
2. Install A8. Remo
ve the jump er from A8TP2 (A
GND) to c
hassis ground.
3. Remove or disconnect all the assemblies listed b elow. See Figure 5-5 .
Always switch o
the analyzer beforeremoving or disconnecting an assembly.
A10 digital IF
A11 phase lo ck
A12 reference
A13 fractional-N analog
DRAFT
3/21/106 15:12
Power Supply Troubleshooting 5-11
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 8753D Option 011 and is the answer not in the manual?

HP 8753D Option 011 Specifications

General IconGeneral
BrandHP
Model8753D Option 011
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals