EasyManuals Logo
Home>HP>Measuring Instruments>8753D Option 011

HP 8753D Option 011 User Manual

HP 8753D Option 011
398 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #73 background imageLoading...
Page #73 background image
Analog
Bus Correction
Constant (T
est
46)
Analyzer
warm-up time:
30
min
utes.
This
pro cedure
calibrates
the
analog
bus
b
y
using
three
reference
v
oltages:
ground, +0.37
V
and
+2.5
V.
The
calibration
data is
stored as
correction constan
ts in
EEPROMs.
1. Press
4
PRESET
54
SYSTEM
5
N
N
N
N
N
N
NN
NN
NN
NN
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
NN
NN
NN
N
SERVICE MENU
N
N
N
N
N
N
NN
NN
NN
NN
N
N
N
TESTS
4
46
54
x1
5
N
N
N
N
N
N
NN
NN
NN
NN
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
NN
NN
NN
N
EXECUTE TEST
N
N
N
N
N
N
NN
NN
N
YES
.
2.
Observ
e
the
analyzer
for
the
results
of
the
adjustment
routine:
If
the analyzer
displays
ABUS Cor
DONE
,
y
ou
ha
v
e
completed
this
pro
cedure.
If
the
analyzer
displa
ys
ABUS
Cor
FAIL
,
refer
to
the
\Digital
Control
Troublesho
oting"
c
hapter.
3-8 Adjustments and Correction Constants DRAFT
3/21/106 15:11
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 8753D Option 011 and is the answer not in the manual?

HP 8753D Option 011 Specifications

General IconGeneral
BrandHP
Model8753D Option 011
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals