A2
Front P
anel Interface
T
ools Required
T-10
TORX
screwdriv
er
T-15
TORX screwdriv
er
small
slot
screwdriv
er
ESD
(electrostatic
discharge)
grounding wrist
strap
5/16-inc
h
op
en-end
torque
wrenc
h
(set
to
10
in-lb)
Remov
al
1.
Remo
v
e
the
front
panel (refer
to \F
ron
t
P
anel
Assem
bly"
in
this
c
hapter).
2.
Disconnect
W18,
W19, and
RPG1W1
from
the
in
terface
b
oard
(A2).
3.
Disconnect
A1W1
from
the
in
terface
b
oard
(A2),
inserting
the
blade
of
a
small
slot
screwdriv
er in
to the
slots
on
the
sides
of
the
ribb
on
cable
connector.
Gen
tly
pry
up
w
ard
on
either
side
of
the
connector
un
til the
ribb on
cable is
detached.
4.
Remo
v
e
the
four
screws
(item
1)
from
the
corners
of
the
in
terface b
oard (A2).
Remo
v
e
the
b
oard.
Replacement
Reverse
the
order
of
the
remo
v
al
pro
cedure.
14-22 Assembly Replacement and
Post-Repair Procedures
DRAFT
3/21/106 15:15
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 8753D Option 011 and is the answer not in the manual?

HP 8753D Option 011 Specifications

General IconGeneral
BrandHP
Model8753D Option 011
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals