Basic Operation
Invert: Yes
Not-Invert: No
Invert
Sync-
Off
This default function means that pin 7
performs bi-directional and synchronous
control over the turn-off of [On/Off].
Input Pin 7 receives the level signal from the
outside.
Output Pin 7 sends the digital signal (1, 0,
PWM) to the outside.
True By default (Not-Invert), the out-
put digital signal is 1 (i.e. low
level), and in the case of
Invert, the output is high level.
False By default (Not-Invert), the out-
put digital signal is 0 (i.e. high
level), and in the case of
Invert, the output is low level.
PWM Digital signal of PWM.
PWM Freq Frequency
PWM Duty Duty cycle
How to Use
When pin 7 is set to default Sync-Off function, pin 7 has bi-directional I/O func-
tion, which can receive pulse signal input from the external instrument and also
can output pulse signal to external instrument. The parameter requirements of
this pulse signal are as follows:
Level rise slope 10us
Level fall slope 2us
Minimum time width
for low level keep
30us
The bi-direction I/O functions are introduced as below:
Copyright © Itech Electronic Co., Ltd.
96

Table of Contents

Other manuals for ITech IT8000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8000 Series and is the answer not in the manual?

ITech IT8000 Series Specifications

General IconGeneral
BrandITech
ModelIT8000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals