Load Function
1. Press the composite keys [Shift]+[V-set] (Config) on the front panel to enter
the configuration menu.
2. Use knob or Up/Down key to select Von and press [Enter].
3. Use knob or left/right key to select the desired Von mode.
• Latch mode
• Living mode
4. Use number keys or knob to set Level and press [Enter].
4.2 Advanced Feature
4.2.1 LIST Function
The IT8000 series load supports a total of 10 List files (List01 to List10), each of
which can be set up to 200 steps. You need to edit the voltage/current value,
slope and time width of each step, or you can set repeat times (1 to 65535) for
each List file. After the file is edited completely, you can trigger the selected List
file to run according to the selected trigger mode.
。The descriptions of list function menu are listed in the table below.
LIST List menu
Run Indicates to enter the List run mode, waiting for the trigger to
run the currently selected List file.
Open Select the List file to open.
USB Open the List file in the external USB flash disk.
Load Import the List file from the exter-
nal USB flash disk into the
instrument.
Not-Load Cancel the import operation.
Internal Open the List file stored inside the instrument.
Recall Inner
List File
Set the name of the recalled List
file.
Export Export the internal List file of the instrument to an external
USB flash drive.
File
Name
The name of the exported list file. The default suf-
fix is .csv, and the file name can only be a string
of numbers.
Copyright © Itech Electronic Co., Ltd.
53