Technical Specification
6.1.4 IT8018-1500-30
Parameters IT8018-1500-30
Rated value ( 0
-40 )
Input voltage
01500V
Input current
030A
Input power
018000W
Input resistance
050000Ω
Min. operating
voltage
9V at 30A
Setup Resolution
Voltage
0.1V
Current 0.001A
Power 0.1W
Resistance 0.1Ω
Read Back
Resolution
Voltage
0.1V
Current 0.001A
Power 0.1W
Resistance 0.1Ω
Setup Accuracy
(within 12 months,
25 ±5) ±(of
Output+Offset)
Voltage
≤0.1% + 1500mV
Current ≤0.1% + 30mA
Power ≤1%Pmax
Resistance
≤2%Rmax,010%Rmax;≤5%Rmax,10%Rmax;
Read Back Accu-
racy (within 12
months, 25 ±5)
±(of Output
+Offset)
Voltage
≤0.1% + 1500mV
Current ≤0.1% + 30mA
Power ≤1%FS
Resistance
≤2%Rmax,010%Rmax;≤5%Rmax,10%Rmax;
Ripple (20Hz
-20MHz)
Voltage ≤1500mVpp
Current ≤30mArms
Setup Temperature
Coefficient (of
Output/+Offset)
Voltage
≤0.01% + 150mV
Current ≤0.02% + 6mA
Read Back Temper-
ature Coefficient (
of Output/+Offset)
Voltage
≤0.01% + 150mV
Current ≤0.02% + 6mA
Dynamic
Rising slope
30A/ms
Falling slope
30A/ms
Dynamic
Frequency
500Hz
Copyright © Itech Electronic Co., Ltd.
118

Table of Contents

Other manuals for ITech IT8000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8000 Series and is the answer not in the manual?

ITech IT8000 Series Specifications

General IconGeneral
BrandITech
ModelIT8000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals