Remote Control 
  Copyright © ITECH Electronic Co., Ltd.  7 
10: 
TRIG:COUN 10 
Then, you may query count value by sending the following command: 
TRIG:COUN? 
You may also query the allowable minimum or maximum count as follows: 
TRIG:COUN?MIN 
TRIG:COUN?MAX 
Comma (,) 
If a command requires several Parameter, then a comma must be used to 
separate adjacent Parameter. 
Space 
You must use blank characters, [TAB] or [Space] to separate Parameter with 
key words of commands. 
Generic commands (*) 
Execute functions like reset, self inspection and status operation. Generic 
commands always start with a asterisk (*) and occupy 3 character sizes, 
including one or more Parameter. Key words of a command and the first 
parameter are separated by a space. Semicolon (;) can separate several 
commands as follows: 
*RST; *CLS; *ESE 32; *OPC? 
Command terminator 
Command strings sent to the instrument must end with a <Newline> (<NL>) 
character. IEEE-488 EOI (End or Identify) information can be used as <NL> 
character to replace termination command string of <NL> character. It is 
acceptable to place one <NL> after a <Enter>. Termination of command string 
always resets current SCPI command path to root level. 
Note 
As for every SCPI message with one query sent to the instrument, the instrument will 
use a <NL> or newline sign (EOI) to terminate response of return. For example, if 
"DISP:TEXT?" is sent, <NL> will be placed after the returned data string to terminate 
response.  If  an  SCPI  message  includes  several  queries  separated  by  semicolon 
(such  as  "DISP?;DISP:TEXT?"),  <NL>  will  terminate  response  returned  after 
response to the last query. In all cases, the program must read <NL> in response 
before another command is sent to the instrument, otherwise errors will be caused. 
1.7 Data Type 
SCPI language defines several data types used for program message and 
response messages. 
⚫  Numerical parameter 
Commands requiring numerical Parameter support the notations of all common 
decimal notations, including optional signs, decimal points, scientific notation, 
etc. Special values of numerical Parameter are also acceptable, such as MIN, 
MAX and DEF. In addition, suffixes for engineering units can also be sent 
together with numerical Parameter (including M, k, m or u). If the command 
accepts only some specific values, the instrument will automatically round the 
input Parameter to acceptable values. The following commands require 
numerical Parameter of frequency value: 
[SOURce[1|2]:]FREQuency:CENTer {<Frequency>|MINimum|MAXimum}