Input Settings
Copyright © ITECH Electronic Co., Ltd. 29
Examples
CURR:PROT 3
Query Syntax
[SOURce:]CURRent:PROTection[:LEVel]?
Returned Parameters
NR2
[SOURce:]CURRent:PROTection[:LEVel]?
This command is used to query the current protection level.
Command Syntax
[SOURce:]CURRent:PROTection[:LEVel]?
Unit
A
*RST Value
MAXimum
[SOURce:]POWer:PROTection[:LEVel] <NRf+>
This command sets the power protection level. If the input power exceeds the
power protection level for the time specified by POW:PROT:DEL, the input is
turned off.
NOTE: Use POW:PROT:DEL to prevent momentary power limit conditions
caused by programmed changes from tripping the over power protection.
Command Syntax
[SOURce:]POWer:PROTection[:LEVel] <NRf+>
Parameters
MIN ~ MAX | MINimum | MAXimum
Unit
W
*RST Value
MAXimum(high range)
Examples
POW:PROT 100