Input Settings
Copyright © ITECH Electronic Co., Ltd. 24
*RST Value
MAXimum (high range)
Examples
CURR:RANGE MIN
Query Syntax
[SOURce:]CURRent:RANGe?
Returned Parameters
<NR2>
[SOURce:]CURRent:RANGe?
This command is used to query the current range of the electronic load module.
Command Syntax
[SOURce:]CURRent:RANGe?
Parameters
None
Unit
A
*RST Value
MAXimum (high range)
[SOURce:]VOLTage:RANGe <NRf+>
This command sets the voltage range of the electronic load module. When you
program a range value, the load automatically selects the range that
corresponds to the value that you program. If the value falls in a region where
ranges overlap, the load selects the range with the highest resolution.
Command Syntax
[SOURce:]VOLTage:RANGe <NRf+>
Parameters
MIN ~MAX | MINimum | MAXimum
Unit
V
*RST Value
MAXimum(high range)