Essential Commands
Copyright © ITECH Electronic Co., Ltd. 16
Parameters
0|1|OFF|ON
Examples
SYST:SENS ON
Query Syntax
SYSTem:SENSe[:STATe]?
Returned Parameters
0|1
*RST Value
OFF
SYSTem:SENSe[:STATe]?
This command is used to query the status of remote compensation.
Command Syntax
SYSTem:SENSe[:STATe]?
Parameters
Bool
Returned Parameters
<NR1>
SYSTem:LOCal
The command sets the load to local mode when in RS232 mode. Can use the
buttons on front panel to operate.
Command Syntax
SYSTem:LOCal
Parameters
None
Examples
SYST:LOC
SYSTem:REMote
The command sets the load to remote mode when in RS232 mode. All the
buttons except for the LOCAL and Shift buttons will lose function. In the remote
state, press LOCAL key return to local mode.