Input Settings
Copyright © ITECH Electronic Co., Ltd. 28
Examples
CURR:SLEW FALL 3
Query Syntax
[SOURce:]CURRent:SLEW:FALL?
Returned Parameters
<NR2>
[SOURce:]CURRent:SLEW:FALL?
This command is used to query the slew rate of the current for negative going
transitions.
Command Syntax
[SOURce:]CURRent:SLEW:FALL?
Parameters
None
Unit
A/uS
*RST Value
MAXimum
[SOURce:]CURRent:PROTection[:LEVel] <NRf+>
This command sets the current protection level. If the input current exceeds the
current protection level for the time specified by CURR:PROT:DEL, the input is
turned off.
NOTE: Use CURR:PROT:DEL to prevent momentary current limit conditions
caused by programmed changes from tripping the overcurrent protection.
Command Syntax
[SOURce:]CURRent:PROTection[:LEVel] <NRf+>
Parameters
MIN ~ MAX | MINimum | MAXimum
Unit
A
*RST Value
MAXimum