EasyManuals Logo
Home>ITech>Test Equipment>IT8731

ITech IT8731 Programming Guide

ITech IT8731
133 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #83 background imageLoading...
Page #83 background image
SOURce Subsystem
Copyright © Itech Electronic Co., Ltd. 71
[SOURce:]TRANsient
This command turns the transient generator on or off.
Command Syntax
[SOURce:]TRANsient[:STATe] <bool>
Parameters
0 | 1 | OFF | ON
*RST Value
OFF
Examples
TRAN 1
Query Syntax
[SOURce:]TRANsient[:STATe]?
Returned Parameters
0 | 1
Related Commands
CURR:TRAN:CURR:MODE CURR:TRAN:ALEV
[SOURce:]INPut:PROTection:CLEar
This command clears the latch that disables the input when a protection
condition such as overvoltage (OV) or overcurrent (OC) is detected. All
conditions that generated the fault must be removed before the latch can be
cleared. The input is then restored to the state it was in before the fault
condition occurred.
Command Syntax
[SOURce:]INPut:PROTection:CLEar
Parameters
None
Examples
INP:PROT:CLE

Table of Contents

Other manuals for ITech IT8731

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITech IT8731 and is the answer not in the manual?

ITech IT8731 Specifications

General IconGeneral
BrandITech
ModelIT8731
CategoryTest Equipment
LanguageEnglish

Related product manuals