5.2 Various problems and countermeasures
If the following problems occur with any of these devices, consult the respective references for countermeasures.
IMPORTANT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
After performing countermeasures, if the problem does not go away, contact Konica Minolta technical representatives.
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
HINT
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
When an error message has been displayed in the image processing controller, check the error description and countermea-
sures listed in the "Operation Manual" of the image processing controller.
When using a general-purpose hub or access point, refer to its operation manual.
•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••
5.2.1 DR Detector
Status Error description Corrective actions
The DR Detector does not start
up.
Power does not go on even when the power
switch is pressed for 2 seconds or longer.
The battery might be out. Charge it as follows.
Then, start it.
Inserting it into the AeroDR Battery Charger
*1
for more than 2 minutes
Inserting it into the AeroDR Battery Charger2
for more than 6 minutes
Connecting a wired cable for more than 6
minutes
The DR Detector does not shut
down.
Power does not go o󰀨 even when the power
switch is pressed for 5 seconds or longer.
It is not possible to shut down while exposing.
Shut down after exposure is ended.
For AeroDR 1417HQ, AeroDR
1417S, AeroDR 1717HQ, AeroDR
1012HQ, AeroDR 2 1417HQ and
AeroDR 2 1417S
The status LED (green) is lit,
and the busy/error LED (orange)
ashes rapidly.
("Ready" is not displayed on the
image processing controller)
System error is occurring.
If an error is displayed for about 10 minutes or
more, shutdown the DR Detector.
Or, when "Ready" is not displayed on the im-
age processing controller, restart it.
For AeroDR 3 1417HD, AeroDR 3
1717HD and AeroDR 3 1012HQ
The Information LED (white)
ashes fast displaying the
"E" mark and the Mode LED
(orange) and the Status LED
(orange) light.
("Ready" is not displayed on the
image processing controller)
When the DR Detector is placed
on a smooth surface, the DR
Detector is not stably attached to
the smooth surface.
DR Detector is warped.
If the DR Detector is still warped even
after the detector is placed on a smooth
surface, contact Konica Minolta technical
representatives.
For AeroDR 1417HQ, AeroDR 1417S, AeroDR
1717HQ, AeroDR 1012HQ, AeroDR 2 1417HQ
and AeroDR 2 1417S
The protective cover is deformed.
Contact Konica Minolta technical
representatives.
For AeroDR 3 1417HD, AeroDR 3 1717HD
and AeroDR 3 1012HQ
The back board is deformed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta SKR 3000 and is the answer not in the manual?

Konica Minolta SKR 3000 Specifications

General IconGeneral
BrandKonica Minolta
ModelSKR 3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals