1.3 Safety precautions
22
1.3.3 Precautions regarding
electromagnetic waves
EMC Statement
This device has been tested and found to comply with
the limits for medical devices in IEC 60601-1-2: 2007
or IEC 60601-1-2: 2014.
These limits are designed to provide reasonable pro-
tection against harmful interference in a typical medi-
cal installation. The device generates, uses and can
radiate radio frequency energy and, if not installed and
used in accordance with the instructions, may cause
harmful interference to other devices in its vicinity.
However, there is no guarantee that interference will
not occur in a particular installation.
Whether this device does cause harmful interference
to other devices can be determined by turning this
device o󰀨 and on. If it causes harmful interference, the
user is encouraged to try to correct the interference by
1 or more of the following measures:
Reorient or relocate the receiving device.
Increase the separation between the devices.
Connect this device into a wall outlet on a circuit di󰀨er-
ent from that to which the other devices are connected.
Contact Konica Minolta technical representatives.

WARNING
Do not use mobile RF communication devices such as
mobile phones or pocket pagers in the vicinity of this
device. Use of such RF devices near this device can
cause errors in operation due to electromagnetic wave
interference. When using a mobile RF device, ensure
that the distance between the RF device in use and this
device is at least 30 cm (12 inchies).
The use of accessories, transducers and cables other
than those sold by Konica Minolta, Inc. as internal
components, may result in increased emissions or de-
creased electromagnetic immunity of this device, pos-
sibly causing malfunction.
Do not use this device adjacent to or stacked with other
devices. If adjacent or stacked use is necessary, con-
rm normal operation in the conguration in which this
device will be used.
Supplementary information regarding IEC 60601-
1-2:2007 and IEC 60601-1-2:2014
(1) Take precautions against this device especially re-
garding EMC. Install and put into service according
to the electromagnetic compatibility (EMC) informa-
tion provided in the manual (Table 1 - Table 5).
(2) This device is suitable for use in medical institutions
such as hospitals and clinics. However, this device is
not suitable for use in environments with high degree
of electromagnetic wave interference (such as near an
electric operation device in use for output, or near a
high frequency treatment device).
(3) Cable list
Power cable (3.048 m/3-Wire/No Shielding)
Ethernet cable (max 30 m/Shielding)
Non-shielded Ethernet cable can be used for con-
necting to the AeroDR Battery Charger and the
AeroDR Battery Charger2
(4) Specications regarding RF transmitters frequency:
Frequency
AeroDR SYSTEM:
5150 MHz to 5350 MHz, 5470 MHz to
5850 MHz
AeroDR SYSTEM 2:
2412 MHz to 2472 MHz
5180 MHz to 5320 MHz, 5500 MHz to
5825 MHz
SKR 3000:
2412 MHz to 2472 MHz
5180 MHz to 5320 MHz, 5500 MHz to
5825 MHz
Modulation
AeroDR SYSTEM:
OFDM
AeroDR SYSTEM 2:
2412 MHz to 2472 MHz: DSSS/CCK/OFDM
5180 MHz to 5320 MHz, 5500 MHz to
5825 MHz:OFDM
SKR 3000:
2412 MHz to 2472 MHz: DSSS/CCK/OFDM
5180 MHz to 5320 MHz, 5500 MHz to
5825 MHz:OFDM
Maximum e󰀨ective radiation power
AeroDR SYSTEM:
+15 dBm
AeroDR SYSTEM 2:
+10 dBm
SKR 3000:
+10 dBm
This device may be interfered with by other devic-
es that conform to CISPR emission requirements.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta SKR 3000 and is the answer not in the manual?

Konica Minolta SKR 3000 Specifications

General IconGeneral
BrandKonica Minolta
ModelSKR 3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals