810
Triggers and When to Use Them
SMART Trigger
SMART Trigger allows the setting of additional qualifications
before a trigger is generated. Depending on the oscilloscope
model, this can include triggers adapted for glitches, intervals,
abnormal signals, TV signals, state- or edge-qualified events,
dropouts and patterns.
Glitch Trigger Glitch Trigger (Fig. 8–6) is used to capture narrow pulses
inferior to or exceeding a given time limit. In addition, a width
range can be defined to capture any pulse that is comprised
within or outside the specified range an Exclusion Trigger.
Figure 8–6. Glitch Trigger Menu (see page 8–30).
Glitch
Trigger on
Coupling
At end of Neg|Pos|Pulse
DC|AC|LFREJ|HFREJ|HF
1|2|3|4|Ext|Ext10|Pattern
Width: l
ess
than or equal
to
Off|Time (2.5 ns to 20 s)
Width:
greater than
or equal to
Off|Time (2.5 ns to 20 s)