D9U001MC0-0101_16
11
MAINS POWER LED MEANING
lit connected to the mains
ashing: 0,6s lit / 0,6s not lit keyboard error (ashing inverted to Lock LED)
error (rst fault)
ashing: 0,1s lit / 0,1s not lit service mode
not lit disconnected from the mains power
transformer switching error
1.8 Visual signalisation
ACCUMULATOR INDICATOR MEANING
lit accumulator disconnected or faulty
ashing: 1,6s lit / 0,2s not lit accumulator deeply discharged
ashing: 0,1s lit / 0,1s not lit accumulator discharged
ashing: 0,2s lit / 1,6s not lit accumulator is charging
not lit accumulator charged
1.8.1 Mains Power LED (Multiboard, Attendant Control Panel)
1.8.2 Accumulator Indicator (Multiboard, Attendant Control Panel)
VISUAL
SIGNALISATION
LOCK LED
lit
ashing:
0,6s lit / 0,6s not lit not lit
Thighrest, Calfrest
and Bed Extension Lock LED
locked lock
error
keyboard
error
motion
blocked
unlocked
Backrest Lock LED locked lock
error
keyboard
error
motion
blocked
unlocked
Bed Height, Lateral Tilt,
Trendelenburg and Antitrendelen-
burg Tilt Lock LED
locked lock
error
keyboard
error
motion
blocked
unlocked
Foot Switch Lock
LED
locked lock
error
keyboard
error
motion
blocked
unlocked
1.8.3 Lock LED (Attendant Control Panel)