D9U001MC0-0101_16
49
Displaying ALT history:
Press ALT history Icon.
Fig. ALT history screen
13 Scales Control
13.1 Scales Screen (WS17)
Fig. Scales Screen
1. History Subscreen Icon
2. Stabilized Scales Status Icon
3. Display – Current Weight
4. Save Weight Value Icon
5. Indicated Value Switch Icon
6. ZERO/T Icon (zero or tare scales)
7. HOLD Icon
8. WEIGHT/CLEAR icon
13.1.1 Preparation
Install mattress and accessories to prepare bed before patient admission and using the scales.
CAUTION!
Incorrect use of scales due to incomplete preparation!
Before each patient admission tare the scales.
13.1.2 Taring
Taring can be done in range of 5kg to 249,5 kg. Taring is used to set “0” on the display before placing the patient on the bed.
Taring must be done with an unloaded bed with mattress, bed sheets, pillows and necessary accessories, without patient. It is reco-
mmended to position mattress platform about 20 cm above the lowest horizontal position.
To tare weight:
Ensure that nothing and nobody touches the bed except you.
Press and hold icon 6 (Zero/T) until weight value starts to ash.
Release icon 6.
Press icon 6 to conrm taring. “0” is shown on the display.
1
2
3
4
5 6 7 8

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LINET Multicare and is the answer not in the manual?

LINET Multicare Specifications

General IconGeneral
BrandLINET
ModelMulticare
CategoryMedical Equipment
LanguageEnglish

Related product manuals