11.2 Castor Control and Bed Transport
CAUTION!
Material damage due to incorrect transport and involuntary movement!
Prior to transport, ensure that the bed is disconnected from the mains.
Prior to transport, ensure that the auxiliary outlet plug (if available) is disconnected from the mains.
Ensure that the castors are locked prior to assembly, disassembly and maintenance.
Ensure that the castors are locked when the bed is occupied.
Hang the mains cable on the appropriate hook on the bed during transport.
Have the bed transported exclusively by nursing personnel and by at least 2 persons.
11.2.1 Central Castor Control
The bed is equipped with central castor´s control and brake system.
The control levers are located in the four corners of the undercarriage.
11.2.2 Bed transport
Transporting the bed:
Adjust bed height to at least 20 cm (9 in.) below maximum height.
Push bed by handles on head board or foot board.
Castor control lever positions:
1. Forward Movement
The front left castor is locked. The bed moves straight ahead. If the bed is
equipped with a fth castor, this castor determines the direction of move-
ment.
2. Unrestricted Movement
All of the castors are unlocked.
3. Braked
All of the castors are braked.
Fig. Positions of Castor Control Lever