16.6.14 I-Drive Power Maintenance
Periodical maintenance of the i-Drive Power must be done by qualied service technician or authorized service organization at least
once a year.
Service technician must check the following:
accumulator status and eventual replacement of accumulator (after maximum of three years of duty)
gas spring – replace if necessary (after maximum of three years of duty)
i-Drive Power wheel – replace if necessary
lifting mechanism – grease if necessary
cables, control elements – replace if necessary
i-Drive Power function
NOTE To continue maintenance please see chapter Maintenance.
16.7 X-Ray Lung Examination
The backrest of the bed consists of HPL and is x-ray translucent. The
bed is equipped with an x-ray cassette hol- der with 2 U-proles under
the backrest. This design allows taking x-ray images of the patient’s
lungs without moving the patient manually.
16.7.1 Necessary Steps before the Examination
NOTE This procedure is suitable for patients who cannot be moved due to critical conditions (e. g. internal bleeding).
Make sure that patient is in centre of bed.
Make sure that backrest is in lowest position and siderails are folded up.
Pull out x-ray cassette holder.
Insert x-ray cassette (format 43×35 cm (16.93 in. x 13.78 in.)).
Push back x-ray cassette holder with x-ray cassette so that the cassette centre indicator is exactly under the edge of the
mattress platform.
Correct position of x-ray cassette holder using the tooth mechanism so that the upper edge of the x-ray cassette is exactly
under the patient’s shoulder line.
Adjust parameters of the x-ray device.
16.7.2 Examination with C-arm
Backrest and seat of the bed are x-ray translucent. The bed is equipped with a column construction. This design allows C-arm-
-assisted operations (mainly cardiological operations such as temporary external cardiostimulation) without moving the patient. The
x-ray tube of the C-arm is located between the undercarriage and the mattress platform.
Necessary Steps before the Operation
Make sure that backrest is in highest position and siderails are folded up.
Position upper part of C-arm (sensor and indicator) above the patient’s chest.
Fig. X-Ray Lung Examination