D9U001MC0-0101_1648
1. ALT history Icon
2. Cycle Setting – Time Icons
3. Cycle Setting – Angle Icons
4. Cycle Counter
5. TEST Icon (Activation)
6. Backrest Angle Indicator (over 30°)
Fig. ALT Screen - Cycle Values Defi nition and Test
3
2
1
4
5
6
De ning values of ALT cycle:
► Activate touchscreen by pressing GO button.
► Set time value by pressing and holding one of icons 2 until desired time value is reached.
► Set angle value by pressing and holding one of icons 3 until desired angle value is reached.
NOTE It is possible to change the time values in 5-minute steps to up to 30 min.
It is possible to change the angle values continuously up to 30 degrees.
12.2 Test ALT cycle
The ALT test cycle is obligatory and serves to prevent risks such as collisions of moving bed parts, extubation of the patient or dis-
connection of the ventilation circuit or the cannulas.
During the test cycle, the bed goes through every de ned ALT angle and stops at every de ned angle level.
Perform ALT test cycle:
► Press icon 5 to start test cycle.
► Hold icon 5 until test cycle is nished.
Acoustic signal sounds.
Icon Start appears instead of icon Test 5.
12.3 ALT Cycle
After performed test the TEST Icon becomes the Activation Icon to start ALT Cycle.
During an ongoing ALT Cycle the Activation Icon becomes the STOP Icon to stop the ongoing ALT Cycle.
During the ongoing ALT Cycle a countdown to the next ALT movement is displayed on the left from the STOP Icon.
To stop cycle:
► Press STOP Icon.
Cycle is stopped.
ALT countdown:
The ALT function comes with an automatic countdown
of the time remaining until the next lateral bed move-
ment. This is convenient for planning any necessary
activities with the patient.
Fig. ALT Screen – Stop Icon