D9U001MC0-0101_16
71
19.1 Cleaning (Multicare)
Prepare for cleaning as follows:
► Put the mattress platform in the highest position.
► Adjust the back and thighrests so that the reverse sides are accessible.
► Disable the function buttons on the control elements using the Attendant Control Panel.
► Disable the foot controls using the Attendant Control Panel.
► Disconnect the bed from the mains.
► Move the bed to the location where it will be cleaned.
► Lock the brakes on the bed.
19.1.1 Daily Cleaning
Clean the following bed parts:
► All control elements for adjusting the bed
► All handles
□ CPR release handle
► Head Board and Foot Board
► Siderails (in highest position)
► Freely accessible mattress surface
► Mobi-Lift®
► Accessory rails
For safe and gentle cleaning:
► Do not use any strong acids or bases (optimum pH range 6 - 8).
► Exclusively use detergents that are suitable for cleaning medical equipment.
► Do not use abrasive powders, steel wool, or other materials and cleaning agents that might damage the mattress
replacement system.
► Never use any corrosive or caustic detergents.
► Never use detergents that deposit calcium carbonate.
► Never use detergents with solvents that might affect the structure and consistency of the plastics (benzene, toluene,
acetone, etc.).
► Clean electrical components carefully and allow them to dry completely.
► Do not immerse SCU in water or steam-clean it.
► Observe local directives regarding infection control.
► Make sure any cleaning agent used is approved by:
► the facility in which the mattress replacement system is to be used.
► by the environmental protection agency of the country in which the mattress replacement system is to be used.
LINET ® recommends the following cleaning agents:
Parts to be cleaned Cleaning agents
Multicare hospital bed ■ Mikrozid, Terralin Protect, Thermosept (Schülke & Mayr)
■ Bacillol AF, Bacillol Rasant, Dismozon Pur, Microbac Forte, Neodisher
Dekonta (BODE Chemie)
■ Lysoformin 3000, Lysoform Spezial (LYSOFORM)
■ Incidin plus, Incidin rapid (Ecolab)
■ Perform, TPH protect (Schülke)
Mattress cover base, comforter covers, air
cells, foam base, SCU
■ standard hospital detergents
■ alcohol or chlor based desinfections
Mattress cover top ■ standard hospital detergents
■ alcohol and quaternary ammonium-based disinfectants