19.1 Cleaning (Multicare)
Prepare for cleaning as follows:
Put the mattress platform in the highest position.
Adjust the back and thighrests so that the reverse sides are accessible.
Disable the function buttons on the control elements using the Attendant Control Panel.
Disable the foot controls using the Attendant Control Panel.
Disconnect the bed from the mains.
Move the bed to the location where it will be cleaned.
Lock the brakes on the bed.
19.1.1 Daily Cleaning
Clean the following bed parts:
All control elements for adjusting the bed
All handles
 □ CPR release handle
Head Board and Foot Board
Siderails (in highest position)
Freely accessible mattress surface
Mobi-Lift®
Accessory rails
For safe and gentle cleaning:
Do not use any strong acids or bases (optimum pH range 6 - 8).
Exclusively use detergents that are suitable for cleaning medical equipment.
Do not use abrasive powders, steel wool, or other materials and cleaning agents that might damage the mattress
replacement system.
Never use any corrosive or caustic detergents.
Never use detergents that deposit calcium carbonate.
Never use detergents with solvents that might affect the structure and consistency of the plastics (benzene, toluene,
acetone, etc.).
Clean electrical components carefully and allow them to dry completely.
Do not immerse SCU in water or steam-clean it.
Observe local directives regarding infection control.
Make sure any cleaning agent used is approved by:
the facility in which the mattress replacement system is to be used.
by the environmental protection agency of the country in which the mattress replacement system is to be used.
LINET ® recommends the following cleaning agents:
Parts to be cleaned Cleaning agents
Multicare hospital bed Mikrozid, Terralin Protect, Thermosept (Schülke & Mayr)
Bacillol AF, Bacillol Rasant, Dismozon Pur, Microbac Forte, Neodisher
Dekonta (BODE Chemie)
Lysoformin 3000, Lysoform Spezial (LYSOFORM)
Incidin plus, Incidin rapid (Ecolab)
Perform, TPH protect (Schülke)
Mattress cover base, comforter covers, air
cells, foam base, SCU
standard hospital detergents
alcohol or chlor based desinfections
Mattress cover top standard hospital detergents
alcohol and quaternary ammonium-based disinfectants