Positioning blocked to avoid collision of
the bed with oor or collision of the bed
with bed equipments.
To continue in positioning, adjust the bed so that there
is no collision.
Horizontal position was reached during
tilting.
Press corresponding button to continue in positioning.
Lateral Tilt stopped in 15° when Patient
Transfer Mode is activated.
For information only.
Maximum Lateral tilt was adjusted by
foot control (15°).
For information only.
Lateral Tilt limited to 15 degrees when
Backrest Angle is in 30 degrees or more.
Lower the Backrest to continue in Lateral Tilting.
Safe Working Load exceeded (more
than 10 kg over Safe Working Load).
Remove load!
Maximum Lateral Tilt 15 degrees (Load
more than 150 kg).
Remove load to enable Lateral Tilting again!
Lateral Tilt disabled (Load more than 200
kg).
Remove load to enable Lateral Tilting again!
System Control Unit (OptiCare or Sym-
bioso) disconnected.
Check that bed is connected to the mains power and
contact service department approved by manufacturer
if the bed with Symbioso integrated mattress is conne-
cted to the mains power and this pop-up remains.
The mattress de ating failed.
Use manual CPR! (System Control Unit of the integra-
ted mattress is disconnected.)
System Fatal Error.
Contact service department approved by manufactu-
rer.
Overwrite con rmation / Con rmation of
changes to settings.
Select tick () for “yes” or cross () for “no”.
Scale module disconnected and Bed Exit
monitoring disabled.
Contact service department approved by manufactu-
rer.