14 - 8 Accutorr 7 Operator’s Manual
14.3.2 Selecting a CMS
If [Select CMS] is enabled, you can select the CMS for the current monitoring.
To select the CMS, select the prompt message area at the bottom of the screen. Then the
selected CMS name will display.
If the CMS you select does not have a name, this area displays “???”.
14.3.3 Clearing the Selected CMS at Startup
You can clear the selected CMS each time the monitor restarts after being powered off
for more than 2 minutes.
To clear the selected CMS,
1. Select [Main] → [Maintenance>>] → [User Settings >>] → enter the required
password → Select [Ok] → [Network >>].
2. Set [Clear CMS IP at startup] to [On].
The selected CMS will not be cleared when only one CMS is configured, or the monitor is
restarted within 2 minutes.
This function is switched off by default.
14.4 Managing Certificates
You can import or delete the monitor’s certificates.
1. Select [Main] → [Maintenance>>] → [User Settings>>] → enter the required
password → [Network >>] → [Certificates Maintenance >>].
2. Select [Import certificates >>] or [Delete certificates >>].
14.5 Setting the Multicast Parameters
To set the Multicast parameters:
1Select [Main] → [Maintenance >>] → [User Settings >>] → Enter the
required password→Select [Ok] to access [User Settings] menu.
2. Select [Network >>]
→ [Multicast Setup >>].
3. Set [Multicast Addr] and [TTL].
14.5.1 DIAP Communication Setup
1. Select [Main] → [Maintenance >>] → [User Settings >>] → Enter the required
password → Select [Ok] → [Others >>]
2. Select [DIAP Communication>>] to access [DIAP Communication Setup] menu.
3. Set the baud rate.
4. Select [Ok] to exit the menu.