Accutorr 7 Operator’s Manual 3 - 7
3.6 Changing General Settings
This section covers only general settings such as language, brightness, date and time,
etc. Measurement settings and other settings can be referred to in the respective
sections.
3.6.1 Setting up a Monitor
To install a monitor or change the monitors location, you need to set it as
follows:
1. Select [Main] [Maintenance >>] [User Settings >>] Enter the required
password Select [Ok] and then access [User Settings] menu.
2. Set up [Monitor Name], [Department] and [Bed No.].
You can set [Changing Bed No.] to:
[Unprotected]: enables you to change Bed No. from the [Patient Demographics]
menu.
[Protected]: prevents you from changing Bed No. from the [Patient
Demographics] menu.
3.6.2 Changing Language
1. Select [Main] [Maintenance >>] [User Settings >>] Enter the required
password Select [Ok] to access [User Settings] menu.
2. Select [Language] and then select the desired language.
3. Restart the monitor.
3.6.3 Configuring the Timeout of Clinician ID
You can configure the retention time of a clinician ID each time it is entered. If a clinician
ID is entered, but there is no activity on the monitor for a configured period of time, the
monitor will clear the ID.
1. Select [Main] [Maintenance >>] [User Settings >>] Enter the required
password Select [Ok] and then access [User Settings] menu.
2. Select [Clinician ID Time out] and then set the time.
3.6.4 Showing/Hiding Patient Name
You can show or hide the current patient name in the main screen. To do so, follow this
procedure:
1. Select [Main] [Maintenance >>] [User Settings >>] Enter the required
password Select [Ok] and then access [User Settings] menu.
2. Select [Patient Demographics].
3. Set [Display Patient Name] to [Ye s ] or [No].
[Ye s ]: the patients name display in the main screen.

Table of Contents

Other manuals for Mindray VS-900c

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray VS-900c and is the answer not in the manual?

Mindray VS-900c Specifications

General IconGeneral
BrandMindray
ModelVS-900c
CategoryMedical Equipment
LanguageEnglish

Related product manuals