D - 10 Accutorr 7 Operator’s Manual
D.2.5 Recorder Alarm Messages
Alarm Message Clearable?
(Yes/No)
Alarm
Level
Cause Solution
Recorder Init
Error
Yes Low An error occurred to
the module, or there
is a problem with
the communication
between the
module and the
monitor.
Restart the monitor.
If the problem still
exists, contact
Mindray or your
service personnel.
Recorder Selftest
Error
Yes Low
Recorder
Unavailable
Yes Low
Recorder Comm
Error
Yes Low
Recorder Vlt High No Low There is a problem
with the system
power supply.
Restart the monitor.
If the message is
prompted for
several times,
contact Mindray or
your service
personnel.
Recorder Vlt Low No Low
Recorder Head
Hot
No Low The recorder has
been working for
too long time.
Stop the recording
and resume the
recording until the
recorder’s print
head cools down.
Rec Head Wrong
Pos.
Yes Low The thermal head of
the recorder is in
wrong position.
Restore the control
lever of the recorder
to its previous
position.
Recorder out of
paper
Yes Low The recorder paper
is used up.
Replace with a new
paper roll.

Table of Contents

Other manuals for Mindray VS-900c

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray VS-900c and is the answer not in the manual?

Mindray VS-900c Specifications

General IconGeneral
BrandMindray
ModelVS-900c
CategoryMedical Equipment
LanguageEnglish

Related product manuals