D - 8 Accutorr 7 Operator’s Manual
D.2.4 CO
2
Alarm Messages
Temp SelfTest Err No High An error occurs
during Temp
module
initialization.
Replace the module.
Temp Over High
Limit
No High The patient’s
temperature is too
high, or an error
occurs.
Reduce the patient’s
temperature, or
replace the module.
Temp Over Low
Limit
No High The patient’s
temperature is too
low, or an error
occurs.
Raise the patient’s
temperature, or
replace the module.
Temp No Probe Yes Med The probe is
disconnected.
Reconnect the
probe.
Temp Probe
Misplaced
Yes Med The temperature
probe is not well
placed, or not
inserted into probe
well.
1. Check if the probe
well is installed.
2. Properly re-insert
the probe into
probe well.
Temp Measuring
Timeout
Yes Med The measuring time
is over 5 minutes in
Monitor mode.
Return the sensor to
the probe well, and
take a temperature
again.
Alarm Message Clearable?
(Yes/No)
Alarm
Level
Cause Solution
Alarm Message Clearable?
(Yes/No)
Alarm
Level
Cause Solution
CO2 Comm Err No High An error occurred
to the CO2
module, or there
is a problem with
the
communication
between the
module and the
monitor.
Restart the
monitor. If the
error remains,
contact Mindray
or your service
personnel.
CO2 Comm
Abnormal
No High
CO2 Comm
Stop
No High
CO2 Init Err No High

Table of Contents

Other manuals for Mindray VS-900c

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray VS-900c and is the answer not in the manual?

Mindray VS-900c Specifications

General IconGeneral
BrandMindray
ModelVS-900c
CategoryMedical Equipment
LanguageEnglish

Related product manuals