A - 4 Accutorr 7 Operator’s Manual
A.5.3 Monitor Interface Specifications
A.5.4 Outputs
A.5.5 Data Storage
A.6 Wireless Network (Optional)
Power 1 AC power input connector
Wired network 1 RJ45 connector
USB 2 standard connectors, USB 2.0
Equipotential Grounding
Terminal
1
Multi-Functions Connector 1 MINI-D RIBBON connector
Nurse Call Signal
Amplitude High level: >3V, providing a maximum of 3 mA output
current;
Low level: <0.5 V, receiving a maximum of 5 mA input
current
Rising and falling time ≤ 1 ms
Alarm Output (Network connector)
Alarm delay time from monitor
to remote equipment
The alarm delay time form the monitor to remote
equipment is ≤2 seconds, measured at the monitor’s
signal output connector.
Trends 5000 groups
Standards IEEE 802.11a/b/g/n
Modulation mode DSSS and OFDM
Operating frequency For 2.4G frequency bands (FCC): 2.4GHz to 2.483GHz
(only use Channels 1 to 11)
For 5G frequency bands (FCC): 5.15GHz to 5.35GHz,
5.725GHz to5.825GHz
Priority All communication data types shall have the same
priority.