Chapter 6 Digital I/O
© National Instruments 6-21 X Series User Manual
DO Pause Trigger Signal
Use the DO Pause Trigger (do/PauseTrigger) signal to mask off samples in
a DAQ sequence. That is, when DO Pause Trigger is active, no samples
occur.
DO Pause Trigger does not stop a sample that is in progress. The pause does
not take effect until the beginning of the next sample.
When you generate digital output signals, the generation pauses as soon as
the pause trigger is asserted. If the source of your sample clock is the
onboard clock, the generation resumes as soon as the pause trigger is
deasserted, as shown in Figure 6-9.
Figure 6-9. DO Pause Trigger with the Onboard Clock Source
If you are using any signal other than the onboard clock as the source of
your sample clock, the generation resumes as soon as the pause trigger is
deasserted and another edge of the sample clock is received, as shown in
Figure 6-10.
Figure 6-10. DO Pause Trigger with Other Signal Source
Pause Trigger
Sample Clock
Pause Trigger
Sample Clock
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments DAQ X Series and is the answer not in the manual?

National Instruments DAQ X Series Specifications

General IconGeneral
BrandNational Instruments
ModelDAQ X Series
CategoryI/O Systems
LanguageEnglish

Related product manuals