Chapter 4 Analog Input
© National Instruments 4-9 X Series User Manual
For example, suppose all channels in a system use a –5 V to 5 V
input range. The signals on channels 0, 2, and 4 vary between
4.3 V and 5 V. The signals on channels 1, 3, and 5 vary between
–4 V and 0 V. Scanning channels in the order 0, 2, 4, 1, 3, 5
produces more accurate results than scanning channels in the
order 0, 1, 2, 3, 4, 5.
4. Avoid Scanning Faster Than Necessary—Designing your system to
scan at slower speeds gives the NI-PGIA more time to settle to a more
accurate level. Here are two examples to consider:
• Example 1—Averaging many AI samples can increase the
accuracy of the reading by decreasing noise effects. In general,
the more points you average, the more accurate the final result.
However, you may choose to decrease the number of points you
average and slow down the scanning rate.
Suppose you want to sample 10 channels over a period of 20 ms
and average the results. You could acquire 500 points from each
channel at a scan rate of 250 kS/s. Another method would be to
acquire 1,000 points from each channel at a scan rate of 500 kS/s.
Both methods take the same amount of time. Doubling the number
of samples averaged (from 500 to 1,000) decreases the effect of
noise by a factor of 1.4 (the square root of 2). However, doubling
the number of samples (in this example) decreases the time the
NI-PGIA has to settle from 4 μs to 2 μs. In some cases, the slower
scan rate system returns more accurate results.
• Example 2—If the time relationship between channels is not
critical, you can sample from the same channel mu
ltiple times and
scan less
frequently. F
or example, suppose an application requires
averaging 100 points from channel 0 and averaging 100 points
from channel 1. You could alternate reading between
channels—that is, read one point from channel 0, then one point
from channel 1, and so on. You also could read all 100 points from
channel 0 then read 100 points from channel 1. The second
method switches between channels much less often and is affected
much less by settling time.
Analog Input Data Acquisition Methods
When performing analog input measurements, you either can perform
software-timed or hardware-timed acquisitions.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments DAQ X Series and is the answer not in the manual?

National Instruments DAQ X Series Specifications

General IconGeneral
BrandNational Instruments
ModelDAQ X Series
CategoryI/O Systems
LanguageEnglish

Related product manuals