Page 102
Auxiliary
P810 Units
This parameter determines whether the Measurement units (P104) are
displayed on the auxiliary line of the display in run mode.
Measurement units will not be displayed
Measurement units will be displayed
P811 Alarms Messages
This parameter determines whether notification messages are displayed on
the auxiliary line of the display in run mode when an alarm relay is switched
on or off. The message is in the form “Alarm High ON”, where the ‘High’ is
determined by the setting of the relay Alarm ID (P212, 222, 232, 242, 252).
Alarm messages will not be displayed
Alarm messages will be displayed
P812 General Control or Pump Messages
This parameter determines whether notification messages are displayed on
the auxiliary line of the display in run mode when a pump or general control
relay is switched on or off. The message is in the form “General 1 ON”,
where the number displayed is the number of the relay.
Pump messages will not be displayed
Pump messages will be displayed
P813 Control Messages
This parameter determines whether notification messages are displayed on
the auxiliary line of the display in run mode when a control relay is switched
on or off. The message is in the form “Time ON”.
Control messages will not be displayed
Control messages will be displayed