• Using and set the tare value,
• Press
,
• Program returns to weighings mode. The inscribed tare value
can be seen on the display with „–” sign,
• Tare can be inscribed anytime in weighings mode.
Notice:
1. You cannot inscribe a new tare value when the tare value in memory
is greater than zero. In the case of trying this the <Err3> message
will be displayed and short audible signal will be emitted.
2. Users can also enter up to 9 tare values to the scale memory
(see 15.10 of his manual).
11.3. Zeroing
To ZERO the scale press: .
The scale will display zero and following pictograms:
and .
Zeroing is only possible within the scope of ±2% of full scale. While
zeroing outside the scope of ±2% you will see <Err2>. Zeroing is
possible only in stable state.
Notice:
Zeroing is possible only within the ±2% interval of the maximal range.
If zeroing is performed beyond this range the <Err2> message and
short audible signal will be emitted.
11.4. Weighings in two ranges
Switching between the I range and the II range happens automatically
(exceeding Max of the I range). Weighings in the second range is
signalled by a pictogram in the top left corner of the display.
Then weighings is done with the accuracy of the II range to the moment
of returning to zero (autozero range
) where the scale switches back
to the I range.