- 73 -
19.6. Continuous transmission
The indicator can work in a continuous transmission mode. It can
be switched on or off in parameters or using RS232 commands.
The frame format sent by the indicator in case of setting
<P2.Prnt> to CntA:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S I space stability space sign mass space Unit CR LF
Stability character
[space] if stable
[?] if not stable
[^] if an indication over the range
[v] if fan indication below the range
sign
[space] for positive values or
[-] for negative values
mass
9 characters justified to the right
unit
3 characters justified to the left
command
3 characters justified to the left
The frame format sent by the indicator in case of setting <P2.Prnt>
to Cntb:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U I stability space sign mass space unit CR LF
19.7. Configuring printouts
General information
If some information included are redundant or not sufficient and there
is a necessity of changes one can design their own protocol format in
EDYTOR WAG computer program. This piece of software is accessible
in: http://www.radwag.com