EasyManuals Logo
Home>RADWAG>Scales>WPT series

RADWAG WPT series User Manual

RADWAG WPT series
85 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #71 background imageLoading...
Page #71 background image
- 71 -
Possible answers:
C0_I CR LF
- command comprehended but cannot be executed
C0_A CR LF
- command comprehended and executed
19.4.10. Switch on continuous transmission in current unit
Syntax: CU1 CR LF
Possible answers:
CU1_I CR LF
- command comprehended but cannot be executed
CU1_A CR LF
MASS FRAME
- command comprehended and in progress
- mass value in current unit is returned
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U I stability space sign mass space unit CR LF
19.4.11. Switch off continuous transmission in current unit
Syntax: CU0 CR LF
Possible answers:
CU0_I CR LF
- command comprehended but cannot be executed
CU0_A CR LF
- command comprehended and executed
19.4.12. Send all implemented commands
Syntax: PC CR LF
Possible answers:
PC_- >_Z,T,TO,S,SI,SU,SUI,C1,C0,CU1,CU0,PC – command executed,
the indicator have sent all the implemented commands.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RADWAG WPT series and is the answer not in the manual?

RADWAG WPT series Specifications

General IconGeneral
BrandRADWAG
ModelWPT series
CategoryScales
LanguageEnglish

Related product manuals