Return to weighing performing the procedure of saving parameters.
Caution:
If the start mass adjustment lasts longer than 15 the <Err8> message
will be displayed and short audible signal will be emitted. Press
to perform calibration again with more stable ambient conditions!
17. COOPERATION WITH PRINTER
Each time the
key is pressed a current mass value together with
mass units is sent to RS 232 interface.
Depending on setting of STAB parameter it can be printed out with
temporary or stable value. Depending on setting of REPL parameter,
printout will be automatic or manual.
One of thermal printer in KAFKA series can cooperate with each
platform scales:
a) KAFKA
Only result of weighing with mass unit can be printed.
b) KAFKA 1/Z
This printer is equipped with an internal real time clock.
Both date and time can be printed.
c) KAFKA SQ S
This printer is equipped with an internal real time clock and
possibility of running statistics from measurements. Statistic
contents: quantity of samples, sum of masses of all samples,
average value, standard deviation, variation factor, min value,
max value, difference max - min.
- 63 -